HR-pQCT
High Resolution peripheral Quantitative CT
Specifications
X-Ray: 68 kVp, 1460 µA, 100 W, Cone beam
Detector Size: 4608 x 400 pixels (221x19.2 mm)
Image Matrix: 512 x 512 - 8192 x 8192 pixels
Bore: Ø 170 mm
Field of View: Ø 140 mm
Scan Length: Max. 200 mm
Voxel Size: Native 30 µm (isotropic)
Resolution: Best < 55 µm (10% MTF)
(Default)
X-Ray: 68 kVp, 1460 µA, 100 W, Cone beam
Integration Time: 43 ms
Projections: 900
Detector Size: 4608 x 400 pixels (221x19.2 mm)
Image Matrix: 2304 x 2304
Bore: Ø 170 mm
Field of View: Ø 140 mm
Voxel Size: 61 µm (isotropic)
Resolution: 95 µm (10% MTF)
One stack: 168 slices / 10.2mm
Scan time: 2 min / one stack
Effective Dose: 〜5 µSv / one stack